Radiation Protection Dosimetry Advance Access published online on July 11, 2008

Radiation Protection Dosimetry, doi:10.1093/rpd/ncn188
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
131/3/399    most recent
ncn188v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Mesbahi, A.
Right arrow Articles by Aslanabadi, N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mesbahi, A.
Right arrow Articles by Aslanabadi, N.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2008. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

COMPARISON OF RADIATION DOSE TO PATIENT AND STAFF FOR TWO INTERVENTIONAL CARDIOLOGY UNITS: A PHANTOM STUDY

Asghar Mesbahi1,*, Parinaz Mehnati1, Ahmad Keshtkar1 and Naser Aslanabadi2

1 Medical Physics Department, Medical School, Tabriz University of Medical Sciences, Attare-Neishabouri Street, Tabriz, Iran
2 Cardiology Department, Medical School, Tabriz University of Medical Sciences, Attare-Neishabouri Street, Tabriz, Iran

* Corresponding author: asgharmesbahi{at}yahoo.com

The purpose of this investigation was to measure the patient and staff dose during routine interventional cardiology procedures for an image intensifier-based and a flat detector system using a water phantom. The Integris BH3000 image intensifier-based (Philips) and the Axiom Artis flat detector-based (Siemens) angiography units were used in this study. The accuracy of tubes potential and irradiation timers and also internal dosimeters were verified and confirmed. A water phantom with a thickness of 18 cm was used for patient and staff dose measurements. For the Philips system, phantom entrance dose rates were 2.77 and 38.97 µGym2 s–1 during fluoroscopy and cineangiography. The respective dose rates for the Siemens were 1.98 and 13.46 µGym2 s–1. Phantom entrance dose rate was 28.5 and 65% higher for the Philips system during fluoroscopy and cineangiography, respectively. Comparing the scattered dose rates at the operator location showed that the flat detector-based Siemens system delivers five times lower dose to the operator in comparison with the image intensifier-based Philips unit. The results suggest that the decrease in received dose of the patient and staff is achievable using the flat detector system. In addition, application of lead curtain and glass is recommended to lower the cardiologist dose especially for the image intensifier-based Philips system.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
A. Mesbahi and N. Aslanabadi
A STUDY ON PATIENTS' RADIATION DOSES FROM INTERVENTIONAL CARDIAC PROCEDURES IN TABRIZ, IRAN
Radiat Prot Dosimetry, December 23, 2008; (2008) ncn310v1.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
A. Mesbahi, N. Aslanabadi, and P. Mehnati
A STUDY ON THE IMPACT OF OPERATOR EXPERIENCE ON THE PATIENT RADIATION EXPOSURE IN CORONARY ANGIOGRAPHY EXAMINATIONS
Radiat Prot Dosimetry, December 16, 2008; (2008) ncn300v1.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.