Radiation Protection Dosimetry Advance Access first published online on June 18, 2006
This version published online on July 25, 2006
Radiation Protection Dosimetry, doi:10.1093/rpd/ncl073
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
SCIENTIFIC AND TECHNICAL PAPER
1 Department of Medical Physics, Salvatore Maugeri Foundation, IRCCS, Scientific Institute of Pavia, via Ferrata 8, 27100 Pavia, Italy
* To whom correspondence should be addressed.
A patient dose survey was carried out measuring the kerma-area product (KAP) values during radiological evaluation in the follow-up of bariatric surgery. The procedures were performed by three radiologists to adjust laparoscopic gastric bands and to detect postoperative complications after Roux-en-Y gastric bypass procedures to treat morbid obesity. Total fluoroscopy time, exposure factors and the overall contribution of fluoroscopy to the accumulated KAP value were recorded. The median KAP values were used to estimate organ doses and effective dose to a standard patient; the radiation risk associated with the procedures was also evaluated. The doses were smaller for one of the three radiologists, owing to a more appropriate beam collimation and a reduction of the screening time. The KAP values ranged from 1.6 to 7.1 Gy cm2 for the laparoscopic adjustable gastric banding management, and from 3.0 and 8.3 Gy cm2 for the radiological examinations after gastric bypass. As a whole, the effective doses associated to these procedures were between 0.5 and 2.7 mSv. The organs receiving the highest doses were not only breast, stomach, pancreas and liver, but also lungs, owing to of their high radiosensitivity, significantly contributed to the effective dose.
Received November 29, 2005
Revised March 10, 2006
Accepted May 7, 2006
PATIENT DOSE DURING RADIOLOGICAL EXAMINATION IN THE FOLLOW-UP OF BARIATRIC SURGERY
Luca Moro 1 *, Claudio Cazzani 2, Orazio Tomarchio 2, Giovanni Morone 3, Antonio Catona 3, and Daniele Fantinato 1
2 Department of Radiology, Salvatore Maugeri Foundation, IRCCS, Scientific Institute of Pavia, via Ferrata 8, 27100 Pavia, Italy
3 Department of Mini-invasive Surgery, Salvatore Maugeri Foundation, IRCCS, Scientific Institute of Pavia, via Ferrata 8, 27100 Pavia, Italy
Luca Moro, E-mail: lmoro{at}fsm.it
![]()
Abstract
This new version changes the order of the authors' names.
![]()
CiteULike
Connotea
Del.icio.us What's this?