Radiation Protection Dosimetry Advance Access first published online on February 3, 2006
This version published online on February 4, 2006
Radiation Protection Dosimetry, doi:10.1093/rpd/nci758
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Leuvens Special Issue Article
1 Medical Physics Unit, ‘Laiko’ Hospital, 17 Ag. Thoma, Goudi, 11527 Athens, Greece
* To whom correspondence should be addressed.
This investigation determined patient doses during digital subtraction angiography (DSA). Fluoroscopy time, dose-area product (DAP) and entrance surface air kerma (ESAK) were analysed from 263 DSA examinations, classified into seven categories: (1) abdominal aorta, iliac, femoral, popliteal and leg arteries; (2) abdominal aorta and superselective DSA of renal arteries; (3) combination of (1) and (2); (4) superselective DSA of common carotid and vertebral arteries, intracranial branches in face and profile projections; (5) superselective DSA of hepatic, splenic, superior and inferior mesenteric arteries; (6) combination of (1) and (4); and (7) celiac trunk and branches. Median DAP values were 67.7, 92.9, 76.6, 53.6, 105.7, 76.1 and 2.6 Gy cm2, respectively. With the exception of one examination, ESAK values were below 2 Gy: the limit for erythema. Compared with published data, DAP values were within the range reported for (1) and (4), slightly larger for (2) and (5), whereas no references were identified for the remaining three categories.
RADIATION DOSES TO PATIENTS FROM DIGITAL SUBTRACTION ANGIOGRAPHY
H. Goni 1,
I. A. Tsalafoutas 2,
G. Tzortzis 3,
P. Pappas 3,
N. Bouzas 3,
J. Loulakas 3,
A. Georgiou 1,
E. Georgiou 4,
and
E. N. Yakoumakis 4 *
2 Medical Physics Unit, ‘Konstantopoulio--Agia Olga’ Hospital, 3-5 Agias Olgas, Nea Ionia, 142 33 Athens, Greece
3 Radiology Department, ‘Laiko’ Hospital, 17 Ag. Thoma, Goudi, 11527 Athens, Greece
4 Medical Physics Department, Medical School, University of Athens, 75 Mikras Asias, 115 27 Athens, Greece
E. N. Yakoumakis, E-mail: egiak{at}cc.uoa.gr
![]()
Abstract ![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
C. J. Francois, D. Tuite, V. Deshpande, R. Jerecic, P. Weale, and J. C. Carr Unenhanced MR Angiography of the Thoracic Aorta: Initial Clinical Evaluation Am. J. Roentgenol., April 1, 2008; 190(4): 902 - 906. [Abstract] [Full Text] [PDF] |
||||