Radiation Protection Dosimetry Advance Access published online on July 20, 2004
Radiation Protection Dosimetry, doi:10.1093/rpd/nch330
Copyright © 2004 Oxford University Press.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
SCIENTIFIC AND TECHNICAL PAPER
1 National Institute of Radiological Sciences, Chiba 263-8555, Japan
* To whom correspondence should be addressed. E-mail: m_takada{at}nirs.go.jp.
The neutron dose equivalent in aircraft is measured using commercial bubble detectors, but at high altitudes there are not only neutrons but also protons. Bubble detectors have been used because they are considered not to be sensitive to protons, but this has been the subject of only a few studies. In this study, by irradiating bubble detectors with energetic protons, the detectors' responses to protons were observed. Bubbles were clearly formed by protons, but the proton sensitivities were one order of magnitude smaller than the neutron sensitivities. Thus, counts of protons can be ignored in neutron measurements in aircraft.
Revised March 4, 2004
Accepted April 28, 2004
MEASURED PROTON SENSITIVITIES OF BUBBLE DETECTORS
2 Cyclotron and Radioisotope Center, Tohoku University, Sendai 980-8579, Japan
![]()
Abstract ![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
R. Machrafi, K. Garrow, H. Ing, M. B. Smith, H. R. Andrews, Yu. Akatov, V. Arkhangelsky, I. Chernykh, V. Mitrikas, V. Petrov, et al. NEUTRON DOSE STUDY WITH BUBBLE DETECTORS ABOARD THE INTERNATIONAL SPACE STATION AS PART OF THE MATROSHKA-R EXPERIMENT Radiat Prot Dosimetry, March 23, 2009; (2009) ncp039v1. [Abstract] [Full Text] [PDF] |
||||