Skip Navigation


Radiation Protection Dosimetry Advance Access originally published online on June 9, 2009
Radiation Protection Dosimetry 2009 135(3):154-161; doi:10.1093/rpd/ncp100
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
135/3/154    most recent
ncp100v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Benmosbah, M.
Right arrow Articles by Asselineau, B.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Benmosbah, M.
Right arrow Articles by Asselineau, B.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2009. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

Energy and fluence calibration of the neutron spectrometer ROSPEC at the IRSN AMANDE facility between 70 keV and 4.5 MeV

M. Benmosbah* and B. Asselineau

Institute of Radioprotection and Nuclear Safety (IRSN), Cadarache, F-13115 Saint-Paul-lez-Durance, France

* Corresponding author: mehdi.benmosbah{at}irsn.fr

Received January 13, 2009, amended April 9, 2009, accepted May 7, 2009

The ROSPEC device is a multi-detector system, which has been designed by Bubble Technologies Industries (BTI at Chalk River, ON, Canada) to assess neutron spectra, and hence neutron dose quantities, at workplace fields. It is made up of six gaseous proportional counters that detect neutrons via the elastic (n,p) scattering (four hydrogenous counters) and with the 3He(n,p)T reaction (two 3He-filled counters). Results of the calibration of a similar rotating spectrometer (ROSPEC) have been described by Rosenstock et al.(1). For energy and fluence calibration purposes, measurements were performed with the accelerator for metrology and neutron applications in external dosimetry (AMANDE) facility at the Laboratory of Neutron Metrology and Dosimetry (Institute of Radiation Protection and Nuclear Safety, IRSN, France). This facility provides monoenergetic neutron radiation fields from 2 keV to 20 MeV. Two kinds of experiments were carried out. First, the ROSPEC was used in its rotational mode for the ISO energies. Then, each detector was irradiated with all the available neutron energies, in a well defined position with the rotation of the device stopped. The energy values of the neutron beam were calculated using the TARGET code. A BC501-A liquid scintillation spectrometer provided the fluence values for energies beyond 1.2 MeV, a methane-filled SP2 counter from 800 keV to 1.4 MeV and an H2-filled SP2 counter from 144 to 800 keV. Reference data for 70 keV monoenergetic neutrons were obtained using the IRSN Long Counter. Results showed that the ROSPEC device was in agreement with the absolute neutron fluences within 10%. Moreover, the new energy calibration factors are in good agreement with those derived by BTI.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.