Radiation Protection Dosimetry Advance Access originally published online on September 13, 2006
Radiation Protection Dosimetry 2007 123(2):262-267; doi:10.1093/rpd/ncl113
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
TECHNICAL NOTES
Radiation exposure and dose evaluation in intraoral dental radiology
1 Medical Radiation Physics Working Group, Carl von Ossietzky University Oldenburg, 26111 Oldenburg, Germany
2 Clinic of Radiotherapy and Oncology, Pius-Hospital, Georgstrasse 12, 26121 Oldenburg, Germany
3 Öko-Institut e.V., Darmstadt Office, Nuclear Engineering & Plant Safety Division, Rheinstrasse 95, 64295 Darmstadt, Germany
* Corresponding author: hui.k.looe{at}mail.uni-oldenburg.de
Received May 20, 2006, amended July 31, 2006, accepted August 8, 2006
| Abstract |
|---|
In this study, dose area product measurements have been performed to propose diagnostic reference levels (DRLs) in intraoral dental radiology. Measurements were carried out at 60 X-ray units for all types of intraoral examinations performed in clinical routine. The third quartile values calculated range from 26.2 to 87.0 mGy cm2. The results showed that there exists a large difference between the patient exposures among different dental facilities. It was also observed that dentists working with faster film type or higher tube voltage are not always associated with lower exposure. The study demonstrated the necessity to have the DRLs laid out as guidelines in dental radiology.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
H. K. Looe, F. Eenboom, N. Chofor, A. Pfaffenberger, M. Steinhoff, A. Ruhmann, A. Poplawski, K. Willborn, and B. Poppe CONVERSION COEFFICIENTS FOR THE ESTIMATION OF EFFECTIVE DOSES IN INTRAORAL AND PANORAMIC DENTAL RADIOLOGY FROM DOSE-AREA PRODUCT VALUES Radiat Prot Dosimetry, June 11, 2008; (2008) ncn172v1. [Abstract] [Full Text] [PDF] |
||||