Skip Navigation


Radiation Protection Dosimetry Advance Access originally published online on February 3, 2006
Radiation Protection Dosimetry 2005 117(1-3):135-138; doi:10.1093/rpd/nci726
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
117/1-3/135    most recent
nci726v2
nci726v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Dierckx, D.
Right arrow Articles by Persyn, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Dierckx, D.
Right arrow Articles by Persyn, K.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2006. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

Patient dosimetry measurements in 50 radiology departments in Belgium

D. Dierckx*, K. Constales, N. Gerardy, T. Goegebuer and K. Persyn

AV Controlatom, Olieslagerslaan, 35 -B-1800 Vilvoorde, Belgium

* Corresponding author: av.controlatom{at}aib-vincotte.be

Since July 2001, the Belgian law requires a yearly quality control measurement on all medical radiological equipment. The medical physicists of the AV Controlatom are certified experts in Belgium and perform these quality control measurements, including a patient dosimetry study. In this study, patient dosimetry results since January 2002 are presented for the following examinations: chest PA and LAT, abdomen and pelvis. Since conventional film-based imaging is currently being replaced by phosphorplate technology, a comparison between both technologies is made. In some radiology departments, digital imaging is already in use. These dosimetry results are also analysed. Dose reductions of >70% can be achieved. For the chest examinations, several departments still use fluoroscopy for positioning the patient. Results show that about half of the patient dose is due to the fluoroscopy. A comparison between the dose results and the European Diagnostic Reference Levels (EDRL) is made.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.