Analysis of neutron and photon response of a TLDALBEDO personal dosemeter on an ISO slab phantom using TRIPOLI-4.3 Monte Carlo code
CEA/Saclay, DEN/DM2S/SERMA/LEPP, Bât. 470, 91191 Gif sur Yvette, France
Corresponding author: yklee{at}cea.fr
TRIPOLI-4.3 Monte Carlo transport code has been used to evaluate the QUADOS (Quality Assurance of Computational Tools for Dosimetry) problem P4, neutron and photon response of an albedo-type thermoluminescence personal dosemeter (TLD) located on an ISO slab phantom. Two enriched 6LiF and two 7LiF TLD chips were used and they were protected, in front or behind, with a boron-loaded dosemeter-holder. Neutron response of the four chips was determined by counting 6Li(n,t)4He events using ENDF/B-VI.4 library and photon response by estimating absorbed dose (MeV g1). Ten neutron energies from thermal to 20 MeV and six photon energies from 33 keV to 1.25 MeV were used to study the energy dependence. The fraction of the neutron and photon response owing to phantom backscatter has also been investigated. Detailed TRIPOLI-4.3 solutions are presented and compared with MCNP-4C calculations.