Can the average glandular dose in routine digital mammography screening be reduced? a pilot study using revised image quality criteria
1 Department of Medical Radiation Physics, Lund University, Malmö University Hospital, SE-205 02 Malmö, Sweden
2 Department of Diagnostic Radiology, Malmö University Hospital, SE-205 02 Malmö, Sweden
3 Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, SE-413 45 Göteborg, Sweden
* Corresponding author: bengt.hemdal{at}rfa.mas.lu.se
There is a need for tools that in a simple way can be used for the evaluation of image quality related to clinical requirements in mammography. The aim of this work was to adjust the present European image quality criteria to be relevant also for digital mammography images, and to use as simple and as few criteria as possible. A pilot evaluation of the new set of criteria was made with mammograms of 28 women from a General Electric Senographe 2000D full-field digital mammography system. One breast was exposed using the standard automatic exposure mode, the other using about half of that absorbed dose. Three experienced radiologists evaluated the images using visual grading analysis technique. The results indicate that the new quality criteria can be used for the evaluation of image quality related to clinical requirements in digital mammography in a simple way. The results also suggest that absorbed doses for the mammography system used may be substantially reduced.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
C. Van Ongeval, A. Van Steen, C. Geniets, F. Dekeyzer, H. Bosmans, and G. Marchal CLINICAL IMAGE QUALITY CRITERIA FOR FULL FIELD DIGITAL MAMMOGRAPHY: A FIRST PRACTICAL APPLICATION Radiat Prot Dosimetry, March 4, 2008; (2008) ncn029v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
T Svahn, B Hemdal, M Ruschin, D P Chakraborty, I Andersson, A Tingberg, and S Mattsson Dose reduction and its influence on diagnostic accuracy and radiation risk in digital mammography: an observer performance study using an anthropomorphic breast phantom Br. J. Radiol., July 1, 2007; 80(955): 557 - 562. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Samei, R. S. Saunders Jr, J. A. Baker, and D. M. Delong Digital Mammography: Effects of Reduced Radiation Dose on Diagnostic Performance Radiology, May 1, 2007; 243(2): 396 - 404. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Bath and L G Mansson Visual grading characteristics (VGC) analysis: a non-parametric rank-invariant statistical method for image quality evaluation Br. J. Radiol., March 1, 2007; 80(951): 169 - 176. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. BRUSIN Digital Mammography: An Update Radiol. Technol., January 1, 2006; 77(3): 226M - 234M. [Abstract] [Full Text] [PDF] |
||||