Radiation Protection Dosimetry 2005 114(1-3):286-293; doi:10.1093/rpd/nch536
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Tingberg, A.
Right arrow Articles by Sjöström, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tingberg, A.
Right arrow Articles by Sjöström, D.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2005. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oupjournals.org

Optimisation of image plate radiography with respect to tube voltage

Anders Tingberg* and David Sjöström

Department of Radiation Physics, Malmö University Hospital, SE-205 02 Malmö, Sweden

* Corresponding author: anders.tingberg{at}rfa.mas.lu.se

Purpose: To find the tube voltage that results in the highest image quality per effective dose unit for chest and pelvis radiography, respectively, using image plates. Methods: Two anthropomorphic phantoms (chest and pelvis) were imaged with several different tube voltages. The mA s settings were chosen so that the effective dose to the phantom was the same, regardless of the tube voltage, for the two examinations, respectively. The quality of the images was evaluated by six experienced radiologists using visual grading analysis. Results: For both the chest and the pelvis examinations, the image quality increased when the tube voltage was reduced compared with the standard settings (125 and 70 kV for chest and pelvis, respectively), which were used for screen-film radiography previously. Conclusions: The image quality of image plate radiography can be increased by lowering the tube voltage compared with the one that was used for screen-film radiography.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Br. J. Radiol.Home page
C S MOORE, J R SAUNDERSON, and A W BEAVIS
Investigating the exposure class of a computed radiography system for optimisation of physical image quality for chest radiography
Br. J. Radiol., September 1, 2009; 82(981): 705 - 710.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
H GEIJER, E NORRMAN, and J PERSLIDEN
Optimizing the tube potential for lumbar spine radiography with a flat-panel digital detector
Br. J. Radiol., January 1, 2009; 82(973): 62 - 68.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.