Skip Navigation

Radiation Protection Dosimetry 2005 114(1-3):269-272; doi:10.1093/rpd/nch546
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Doyle, P.
Right arrow Articles by Gentle, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Doyle, P.
Right arrow Articles by Gentle, D.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2005. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oupjournals.org

Dose–image quality optimisation in digital chest radiography

P. Doyle*, C. J. Martin and D. Gentle

Health Physics, Department of Clinical Physics and Bio-Engineering, Gartnavel Royal Hospital, Glasgow G12 OXH, UK

* Corresponding author: philip.doyle{at}northglasgow.scot.nhs.uk

A range of techniques has been developed in conventional radiography for the chest, because of technical difficulties in imaging the wide variation of tissue densities. These techniques can differ significantly in tube potential selection, for scatter reduction and in application of manual or automatic exposure control. In this study a geometric chest phantom designed to simulate a chest image, has been used to study image quality on an indirect digital radiography (DR) system. Relative values for the signal-to-noise ratio (SNR), have been derived from analysis of contrast detail objects within the lung, heart and subdiaphragm regions of the phantom. Results showing the variation in SNR and detail perception with tube potential are presented for three radiographic techniques: using a grid, an air gap and no scatter reduction. SNR measurements provided a useful objective measure for assessment of image quality for optimising DR systems.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
I. A. Tsalafoutas, G. A. Blastaris, A. S. Moutsatsos, P. S. Chios, and E. P. Efstathopoulos
Correlation of image quality with exposure index and processing protocol in a computed radiography system
Radiat Prot Dosimetry, June 1, 2008; 130(2): 162 - 171.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
K. Kepler and A. Vladimirov
Optimisation strategies introduced for CR at health care centres in Estonia
Radiat Prot Dosimetry, March 1, 2008; 129(1-3): 127 - 131.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.