Radiation Protection Dosimetry 2005 114(1-3):214-219; doi:10.1093/rpd/nch506
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Illers, H.
Right arrow Articles by Neitzel, U.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Illers, H.
Right arrow Articles by Neitzel, U.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2005. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oupjournals.org

Measurement of the modulation transfer function of digital X-ray detectors with an opaque edge-test device

Hartmut Illers1,*, Egbert Buhr1, Susanne Günther-Kohfahl2 and Ulrich Neitzel2

1 Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116 Braunschweig, Germany
2 Philips Medical Systems, Röntgenstrasse 24, D-22335 Hamburg, Germany

* Corresponding author: hartmut.illers{at}gsf.de

A variant of the edge method for the determination of the pre-sampled modulation transfer function (MTF) of digital X-ray imaging devices has been developed and accepted as the standard method in the novel DQE measurement standard IEC 62220-1. An opaque tungsten edge-test device accomplishes the ideal step-like profile of the incident X rays. The edge spread function is measured over a large region across the edge transition that enables an accurate MTF measurement including the ‘low-frequency drop’. The method has been applied to different state-of-the-art X-ray imaging detectors, a computed radiography, a CsI-based indirect and an Se-based direct flat-panel detector. The MTF measurement results will be presented. In contrast to the opaque edge device, the commonly used semi-transparent edge-test devices produce scatter radiation that deteriorates the incident X-ray profile, which leads to a systematic overestimation of the MTF.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
RadioGraphicsHome page
M. Korner, C. H. Weber, S. Wirth, K.-J. Pfeifer, M. F. Reiser, and M. Treitl
Advances in Digital Radiography: Physical Principles and System Overview
RadioGraphics, May 1, 2007; 27(3): 675 - 686.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.