Skip Navigation

Radiation Protection Dosimetry 2004 110(1-4):705-710; doi:10.1093/rpd/nch219
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Buchillier, T.
Right arrow Articles by Wernli, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Buchillier, T.
Right arrow Articles by Wernli, C.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Radiation Protection Dosimetry Vol. 110 Nos. 1-4 © Oxford University Press 2004; all rights reserved

Calibration and testing of a TLD dosemeter for area monitoring

T. Buchillier1,*, A. Aroua1, F. Bochud1, C. Schuler2, N. Stritt3, J.-F. Valley1 and C. Wernli2

1 Institut universitaire de radiophysique appliquée, Grand-Pré 1, CH-1007 Lausanne, Switzerland
2 Paul Scherrer Institute, CH-5232 Villigen PSI, Switzerland
3 Swiss Federal Office of Public Health, CH-3003 Bern, Switzerland

* Corresponding author: Thierry.Buchillier{at}hospvd.ch

The response of a TLD-600/TLD-700 area dosemeter has been characterized in neutron fields around the 590 MeV cyclotron ring at the Paul Scherrer Institute (PSI). The dosemeter is based on a cylindrical paraffin moderator with three of each type of TLD chip at the centre, and is intended to use for area monitoring around accelerator facilities. The dosemeter is calibrated in terms of ambient dose equivalent using a non-moderated 252Cf neutron source. The ambient dose equivalent response has been tested in five locations where the neutron fields and dose rates have been well characterized by Bonner sphere spectrometer and active neutron monitor measurements. The different spectrum shapes and dose rates in the five locations permit the comparison of the behavior of the active and passive dosemeters in these neutron fields.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
C. Wernli
Individual monitoring at accelerator centres
Radiat Prot Dosimetry, November 1, 2009; 137(1-2): 74 - 82.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.