Radiation Protection Dosimetry Vol. 110 Nos. 1-4 © Oxford University Press 2004; all rights reserved
Numerical and experimental results of the operational neutron dosemeter Saphydose-N
1 Institute for Radiological Protection and Nuclear Safety, BP 17, 92262 Fontenay-aux-Roses, France
2 IAEA Grant-holder, DESS Radioprotection Trainee at IRSN
* Corresponding author: thierry.lahaye{at}irsn.fr
Since 1993, the Institute for Radiological Protection and Nuclear Safety (IRSN) has lead, in association with Electricité de France (EDF), a R&D study of a neutron personal electronic dosemeter. This dosemeter, called Saphydose-N, is manufactured by the SAPHYMO company. This paper presents first the optimisation of some detector components using Monte Carlo calculations, and second the test of the manufactured Saphydose-N under radiation following the IEC 1323 standard's recommendations for active personal neutron dosemeters. The measurements with the manufactured dosemeter were performed on the one hand at PTB (Physikalisch-Technische Bundesanstalt) in mono-energetic neutron fields and, on the other hand at IRSN in neutron fields generated by a thermal facility (SIGMA), radionuclide ISO sources and a realistic spectrum (CANEL/T400). The manufactured dosemeter Saphydose-N was also tested during measurement campaigns of the European programme EVIDOS (Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields) at different nuclear workplaces. The study showed that Saphydose-N complies with the recommendations of standard IEC 1323 and can be used at any workplace with no previous knowledge of the neutron field characteristics.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
T. Nunomiya, S. Abe, K. Aoyama, and T. Nakamura DEVELOPMENT OF ADVANCED-TYPE MULTI-FUNCTIONAL ELECTRONIC PERSONAL DOSEMETER Radiat Prot Dosimetry, August 17, 2007; (2007) ncm059v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Luszik-Bhadra Electronic personal neutron dosemeters for high energies: measurements, new developments and further needs Radiat Prot Dosimetry, August 1, 2007; 126(1-4): 487 - 490. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Menard, C. Louis, T. Lahaye, and Q. Chau Simulations of the mean chord length of a multi-element TEPC irradiated by monoenergetic neutrons Radiat Prot Dosimetry, December 20, 2005; 116(1-4): 185 - 189. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bolognese-Milsztajn, M. Ginjaume, M. Luszik-Bhadra, F. Vanhavere, W. Wahl, and A. Weeks Active personal dosemeters for individual monitoring and other new developments Radiat Prot Dosimetry, November 22, 2004; 112(1): 141 - 168. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Luszik-Bhadra Electronic personal dosemeters: the solution to problems of individual monitoring in mixed neutron/photon fields? Radiat Prot Dosimetry, August 1, 2004; 110(1-4): 747 - 752. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bolognese-Milsztajn, D. Bartlett, M. Boschung, M. Coeck, G. Curzio, F. d'Errico, A. Fiechtner, V. Giusti, V. Gressier, J. Kyllonen, et al. Individual neutron monitoring in workplaces with mixed neutron/photon radiation Radiat Prot Dosimetry, August 1, 2004; 110(1-4): 753 - 758. [Abstract] [Full Text] [PDF] |
||||