Radiation Protection Dosimetry Advance Access published online on March 4, 2008

Radiation Protection Dosimetry, doi:10.1093/rpd/ncn047
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
129/1-3/271    most recent
ncn047v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Walsh, C.
Right arrow Articles by Malone, J. F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Walsh, C.
Right arrow Articles by Malone, J. F.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2008. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

QUALITY ASSURANCE OF COMPUTED AND DIGITAL RADIOGRAPHY SYSTEMS

C. Walsh1,2,*, D. Gorman1, P. Byrne1, A. Larkin1, A. Dowling1 and J. F. Malone1

1 Haughton Institute, Dublin 8 & The Adelaide and Meath Hospital incorporating the National Children's Hospital, Dublin
2 Haughton Institute & St. James's Hospital, Dublin 8, Ireland

* Corresponding author: colin.walsh{at}amnch.ie

Computed radiography (CR) and digital radiography (DR) are replacing traditional film screen radiography as hospitals move towards digital imaging and picture archiving and communication systems (PACS). Both IPEM and KCARE have recently published quality assurance and acceptance testing guidelines for DR. In this paper, the performance of a range of CR and DR systems is compared. Six different manufacturers are included. Particular attention is paid to the performance of the systems under automatic exposure control (AEC). The patient is simulated using a range of thicknesses of tissue equivalent material. Image quality assessment was based on detector assessment protocols and includes pixel value measures as well as subjective assessment using Leeds Test Objects. The protocols for detector assessment cover a broad range of tests and in general detectors (whether DR or CR) performed satisfactorily. The chief limitation in performing these tests was that not all systems provided ready access to pixel values. Subjective tests include the use of the Leeds TO20. As part of this work, suggested reference values are provided to calculate the TO20 image quality factor. One consequence of moving from film screen to digital technologies is that the dynamic range of digital detectors is much wider, and increased exposures are no longer evident from changes in image quality. As such, AEC is a key parameter for CR and DR. Dose was measured using a standard phantom as a basic means of comparing systems. In order to assess the AEC performance, exit doses were also measured while varying phantom thickness. Signal-to-noise ratios (SNRs) were calculated on a number of systems where pixel values were available. SNR was affected by the selection of acquisition protocol. Comparisons between different technologies and collation of data will help refine acceptance thresholds and contribute to optimising dose and image quality.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Dentomaxillofac RadiolHome page
D Benchimol, K Nasstrom, and X. Shi
Evaluation of automatic exposure control in a direct digital intraoral system
Dentomaxillofac. Radiol., September 1, 2009; 38(6): 407 - 412.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
J. Zoetelief and K. Faulkner
SAFETY AND EFFICACY FOR NEW TECHNIQUES AND IMAGING USING NEW EQUIPMENT TO SUPPORT EUROPEAN LEGISLATION: AN EU COORDINATION ACTION
Radiat Prot Dosimetry, September 7, 2008; (2008) ncn242v1.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.