Radiation Protection Dosimetry 55:295-298 (1994)
© 1994 Oxford University Press
Inherent Calibration of Microdosemeters for Dose Distribution in Lineal Energy
A method, utilising the inherent electron event spectra, is described for the absolute calibration of microdosemeters in the presence of a photon field. The method, which avoids the problems and uncertainties present in conventional calibration techniques, involves simple extrapolation of the dose distribution in lineal energy associated with 'exact stopper' electrons. Validation of the method is made using the published experimental distributions of Rossi, of Kliauga, and of Dovorak and by direct theoretical calculation of the components of the microdose distributions for gamma rays. Further experimental data from a cylindrical TEPC in a photon field generated by an external source of 137Cs are obtained from comparison. A 'universal' calibration curve for the dose-weighted lineal energy as a function of the simulated mean diameter of the microdosemeter, is presented for use in practical applications.