Radiation Protection Dosimetry 54:189-192 (1994)
© 1994 Oxford University Press
Experiences of Calibration in Photon Beams for the Personal Dose Equivalent
The calibration quantity shall, according to ICRU, be the personal dose equivalent, Hp(d), in a phantom having the composition of the ICRU tissue and the same shape and size as the recommended PMMA calibration phantom, 30 x 30 x 15 cm3. There exist differences in backscattering between PMMA and tissue that for certain photon energies could be of importance. This could either be treated as a systematic uncertainty or be incorporated in the definition. However, monoenergetic beams seldom appear in reality and the difference in backscatter is not thought to be important. The calibration quantity for photons was chosen as the absorbed dose to ICRU tissue (times a quality factor 1) at 10 mm depth in a tissue-equivalent slab phantom. In Sweden 13 different services run personal dosimetry. Their initial hesitation about the change of quantity disappeared after testing their photon energy responses. It was found that most TLD systems were measuring the new quantity better than the old one and that the film systems needed only minor corrections. Most TLD systems now report 5% larger dose equivalents for the same irradiation in a photon beam from a 137Cs source.