Radiation Protection Dosimetry 28:77-81 (1989)
© 1989 Oxford University Press
Determination of Dose Equivalent with Tissue-Equivalent Proportional Counters
Low pressure tissue-equivalent proportional counters (TEPC) are instruments based on the cavity chamber principle and provide spectral information on the energy loss of single charged particles crossing the cavity. Hence such detectors measure absorbed dose or kerma and are able to provide estimates on radiation quality. During recent years TEPC based instruments have been developed for radiation protection applications in photon and neutron fields. This was mainly based on the expectation that the energy dependence of their dose equivalent response is smaller than that of other instruments in use. Recently, such instruments have been investigated by intercomparison measurements in various neutron and photon fields. Although their principles of measurements are more closely related to the definition of dose equivalent quantities than those of other existing dosemeters, there are distinct differences and limitations with respect to the irradiation geometry and the determination of the quality factor. The application of such instruments for measuring ambient dose equivalent is discussed.
![]()