Entrances skin dose distribution maps for interventional neuroradiological procedures: a preliminary study
Struttura Complessa Fisica Sanitaria, Azienda Ospedaliera San Giovanni Battista, Corso Bramante 88, 10126 Torino, Italy
* Corresponding author: orampado{at}molinette.piemonte.it
Dose estimation in interventional neuroradiology can be useful to limit skin radiation injuries. The purpose of this study was to evaluate the role of entrance skin dose (ESD) maps in planning exposure condition optimisation. Thirteen cerebral angiography and five embolisation procedures were monitored, measuring ESD, dosearea product (DAP) and other operational parameters. A transmission ionisation chamber, simultaneously measuring air kerma and DAP, measured dose-related quantities. Data acquisition software collected dosimetric and geometrical data during the interventional procedure and provided a distribution map of ESD on a standard phantom digital image, with maximum value estimation. Values of 881710 mGy for maximum skin dose and 16.7343 Gy cm2 for DAP were found. These data confirm the possibility of deterministic effects during therapeutic interventional neuroradiological procedures like cerebral embolisation. ESD maps are useful to retrospectively study the exposure characteristics of a procedure and plan patient exposure optimisation.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
T. Moritake, Y. Matsumaru, T. Takigawa, K. Nishizawa, A. Matsumura, and K. Tsuboi Dose Measurement on Both Patients and Operators during Neurointerventional Procedures Using Photoluminescence Glass Dosimeters AJNR Am. J. Neuroradiol., November 1, 2008; 29(10): 1910 - 1917. [Abstract] [Full Text] [PDF] |
||||