The performance of commercial photodiodes for dosimetry in mammography
1 Departamento de Energia Nuclear, Universidade Federal de Pernambuco, Cidade Universitária, 50740-540 Recife, PE, Brazil
2 Departamento de Física, Universidade Católica, 50050-900 Recife, PE, Brazil
3 Departamento de Física, Universidade Federal de Pernambuco, Cidade Universitária, 50670-901 Recife, PE, Brazil
* Corresponding author: khoury{at}ufpe.br
This paper reports the results of an investigation carried out to determine the suitability of using a commercially available photodiode for dosimetry in mammography thereby providing a low cost dosemeter with a direct reading of the radiation dose. A mammographic X-ray generator (Siemens Mammomat 1000) with tube potential in the 2330 kV range and a constant potential X-ray generator (Pantak) with a W/Mo anode/filter combination were used in this study. The results showed that the photodiode response is highly linear within mammographic dose ranges and that the energy dependence was <3% at tube potentials of 2530 kV. Good agreement was observed between the incident air kerma measured with both the photodiode and the ionization chamber. These results show the viability of using the photodiode as a dosemeter system in mammography.