Skip Navigation

Radiation Protection Dosimetry 2005 115(1-4):337-339; doi:10.1093/rpd/nci044
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Khoury, H. J.
Right arrow Articles by Lopes, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Khoury, H. J.
Right arrow Articles by Lopes, C.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2005. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

Evaluation of patient dose for mammography in Pernambuco, Brazil

H. J. Khoury*, V. S. Barros and C. Lopes

Departamento de Energia Nuclear—UFPE, Av. Prof. Luiz Freire, 1000, 50740-540 Recife, PE, Brazil

* Corresponding author: khoury{at}ufpe.br

The aim of this study is to evaluate the incident air kerma (INAK) and the mean glandular dose (MGD) during mammography in Recife, North-east Brazil. The study was performed in three institutions: a private clinic (A), a public hospital (B) and a private hospital (C). The incident air kerma to mammography procedures were estimated using a standard breast phantom and the patient MGD, evaluated in the units B and C, was performed using the calibration of output method. This method involves the recording of the patient exposure parameters (kVp, mAs, filter) and the measurement of these parameters of the incident air kerma using a calibrated ionisation chamber. The results obtained showed that INAK for the standard breast phantom, in the institutions A, B and C, were 18.2, 9.4 and 8.5 mGy, respectively. The results of patient MGD values ranged from 0.23 to 7.46 mGy per film, and the MGD values per woman were between 8.23 and 40.6 mGy for the C institution and between 1.50 and 18.78 mGy for the B institution.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
L. C. G. Oliveira, T. K. Dias, R. T. Lopes, and S. Kodlulovich
EVALUATION OF THE ENTRANCE SURFACE AIR KERMA IN MAMMOGRAPHIC EXAMINATIONS IN RIO DE JANEIRO, BRAZIL
Radiat Prot Dosimetry, March 14, 2009; (2009) ncp031v1.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.