Skip Navigation

Radiation Protection Dosimetry 2005 115(1-4):154-158; doi:10.1093/rpd/nci264
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (10)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Polf, J. C.
Right arrow Articles by Titt, U.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Polf, J. C.
Right arrow Articles by Titt, U.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2005. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

Patient neutron dose equivalent exposures outside of the proton therapy treatment field

J. C. Polf*, W. D. Newhauser and U. Titt

Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Boulevard, Unit 130, Houston, TX 77030, USA

* Corresponding author: jpolf{at}yahoo.com

A large fraction of dose to healthy tissue located outside of the treatment field during proton therapy is attributable to neutrons produced in the beam-delivery apparatus. In this work, the neutron dose equivalent (H) per therapeutic proton absorbed dose (D) was estimated for typical treatment conditions as a function of range modulation width, angle with respect to the incident proton beam, and the distance from the isocentre at the Harvard Cyclotron Laboratory's (Cambridge, MA) passively spread treatment field using Monte Carlo simulations. For a beam with 16 cm penetration (depth) and a 5 x 5 cm2 lateral field size at the patient location along the incident beam direction at 100 cm from the isocentre, the predicted H/D values are 0.35 and 0.60 mSv Gy–1 from the simulations and measurements, respectively. At all locations, the predicted H/D values are within a factor of 2 and 3 of the measured result for no modulation and 8.2 cm of modulation, respectively.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
JOURNAL OF THE ICRUHome page
REFERENCES
J. ICRU, December 1, 2007; 7(2): 189 - 210.
[PDF]



Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.