Skip Navigation

Radiation Protection Dosimetry 2005 114(1-3):369-374; doi:10.1093/rpd/nch579
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Klausz, R.
Right arrow Articles by Shramchenko, N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Klausz, R.
Right arrow Articles by Shramchenko, N.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2005. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oupjournals.org

Dose to population as a metric in the design of optimised exposure control in digital mammography

R. Klausz* and N. Shramchenko

Mammography Engineering Department, GE Healthcare, 283 rue de la Minière, 78530 Buc, France

* Corresponding author: Remy.Klausz{at}med.ge.com

This paper describes a method for automatic optimisation of parameters (AOP) in digital mammography systems. Using a model of the image chain, contrast to noise ratio (CNR) and average glandular dose (AGD) are computed for possible X-ray parameters and breast types. The optimisation process consists of the determination of the operating points providing the lowest possible AGD for each CNR level and breast type. The proposed metric for the dose used in the design of an AOP mode is the resulting dose to the population, computed by averaging the AGD values over the distribution of breast types in the population. This method has been applied to the automatic exposure control of new digital mammography equipment. Breast thickness and composition are estimated from a low dose pre-exposure and used to index tables containing sets of optimised operating points. The resulting average dose to the population ranges from a level comparable to state-of-the-art screen/film mammography to a reduction by a factor of two. Using this method, both CNR and dose are kept under control for all breast types, taking into consideration both individual and collective risk.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.