Radiation Protection Dosimetry Advance Access originally published online on July 20, 2004
Radiation Protection Dosimetry 2004 111(3):271-274; doi:10.1093/rpd/nch336
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Radiation Protection Dosimetry Vol. 111, No. 3 © Oxford University Press 2004; all rights reserved
Effective dose to patient during cardiac interventional procedures (Prague workplaces)
Brehova 7, Prague 115 19, Czech Republic
* Corresponding author: stisova{at}hroch.ujf.cao.z
The aim of this study was to assess effective dose to a patient during cardiac procedures, such as coronary angiography (CA) and percutaneous transiluminal angioplasty (PTCA). Measurements were performed on 185 patients in four catheterisation laboratories in three hospitals in Prague using the dose area product (DAP) meter. Calculations of surface and effective dose were performed with Monte-Carlo-based program PCXMC. The mean DAP value per procedure determined in all workplaces ranged between 25.0 and 54.5 Gy cm2 for CA and 43.0104.5 Gy cm2 for PTCA. In three cases, the surface dose exceeded the 2 Gy level for occurrence of transient erythema. The mean effective dose per procedure in all workplaces was determined to be in the range of 2.78.8 mSv for CA and 5.715.3 mSv for CA + PTCA combined. The results presented are comparable with those published by other authors.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
A. Mesbahi and N. Aslanabadi A STUDY ON PATIENTS' RADIATION DOSES FROM INTERVENTIONAL CARDIAC PROCEDURES IN TABRIZ, IRAN Radiat Prot Dosimetry, December 23, 2008; (2008) ncn310v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Bogaert, K. Bacher, and H. Thierens A large-scale multicentre study in Belgium of dose area product values and effective doses in interventional cardiology using contemporary X-ray equipment Radiat Prot Dosimetry, February 1, 2008; 128(3): 312 - 323. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Einstein, K. W. Moser, R. C. Thompson, M. D. Cerqueira, and M. J. Henzlova Radiation Dose to Patients From Cardiac Diagnostic Imaging Circulation, September 11, 2007; 116(11): 1290 - 1305. [Full Text] [PDF] |
||||
![]() |
G. M. Feuchtner, T. Schachner, J. Bonatti, G. J. Friedrich, P. Soegner, A. Klauser, and D. z. Nedden Diagnostic Performance of 64-Slice Computed Tomography in Evaluation of Coronary Artery Bypass Grafts Am. J. Roentgenol., September 1, 2007; 189(3): 574 - 580. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Feuchtner, S. Muller, J. Bonatti, T. Schachner, C. Velik-Salchner, O. Pachinger, and W. Dichtl Sixty-Four Slice CT Evaluation of Aortic Stenosis Using Planimetry of the Aortic Valve Area Am. J. Roentgenol., July 1, 2007; 189(1): 197 - 203. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Bouvier, D. Logeart, J.-L. Sablayrolles, J. Feignoux, C. Scheuble, T. Touche, G. Thabut, and A. Cohen-Solal Diagnosis of aortic valvular stenosis by multislice cardiac computed tomography Eur. Heart J., December 2, 2006; 27(24): 3033 - 3038. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Feuchtner, W. Dichtl, G. J. Friedrich, M. Frick, H. Alber, T. Schachner, J. Bonatti, A. Mallouhi, T. Frede, O. Pachinger, et al. Multislice Computed Tomography for Detection of Patients With Aortic Valve Stenosis and Quantification of Severity J. Am. Coll. Cardiol., April 4, 2006; 47(7): 1410 - 1417. [Abstract] [Full Text] [PDF] |
||||