Skip Navigation

Radiation Protection Dosimetry 2004 111(1):59-64; doi:10.1093/rpd/nch361
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (1)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Campi, F.
Right arrow Articles by Rusconi, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Campi, F.
Right arrow Articles by Rusconi, R.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Radiation Protection Dosimetry Vol. 111, No. 1 © Oxford University Press 2004; all rights reserved

Uncertainty evaluation of radon measurements with LR115 detector and spark counter

F. Campi1, M. Caresana1,*, M. Ferrarini1, L. Garlati1, M. Palermo2 and R. Rusconi2

1 Dipartimento di Ingegneria Nucleare, Politecnico di Milano, via Ponzio 34/3, 20133 Milano, Italy
2 ARPA Lombardia, Dipartimento Città di Milano, via Juvara 22, 20129 Milano, Italy

* Corresponding author: marco.caresana{at}polimi.it

To evaluate the uncertainties for nuclear track detectors used in radon measurements, a full understanding is required of the physical phenomena involved and the behaviour of the instruments utilised in the measuring process. As it concerns the LR115 nuclear track detector, an overall evaluation of uncertainty was given. It was assessed taking into account different contributions and determining their relative weights. Since such detectors are often read by a spark counter device, a model to describe its behaviour was developed and a saturation factor was estimated. Its expression and its associated uncertainty are given. Hence, it has been possible to draw a calibration curve, in which all the uncertainty sources have been considered.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
M. Caresana, F. Campi, and M. Ferrarini
Evaluation of etching correction factor for LR115 cellulose nitrate films from track parameters
Radiat Prot Dosimetry, May 16, 2005; 113(4): 354 - 358.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.